RULLA 2 Wet-Penetration-Test

Harga :
CALL
Pengiriman :
3 hari sejak pembayaran
Lokasi :
Jawa Timur - Surabaya
Dilihat :
373 Kali
Update Terakhir :
12-09-2013 11:45

Dapatkan Spesifikasi Harga RULLA 2 Wet-Penetration-Test Produk Terbaik dari CV. BERKAT JAYA ABADI Berkat Jaya Abadi menyediakan peralatan di bidang Biomolecular, Bahan Kimia, Surface Aerator, Alat-alat Laboratorium.

Detail Produk

l fixed ratio toothed wheel transmission ( no slippage)
l load compensated motor system
l electronic timer
l housing; finger and agar plate holder made of stainless steel

( 1) Counterweight

( 7) Excenter

( 2) Balancing arm with finger

( 8) Turntable

( 3) Dynamometer attachment point

( 9) Timer

( 4) Stainless steel finger

( 10) Ball bearing

( 5) Petri dish ( optional)

( 11) Cover-/ carrier-/ reference material

( 6) Tenter frame/ double steel ring
( optional)

( optional)
INFECTION CONTROL AND HEALTHCARE
Wet-Bacterial-Penetration-Test

Apparatus for measuring the resistance to wet bacterial penetration of surgical drapes and surgical gowns

Test according to EN ISO 22610
Principle of operation is based on and in accordance with EN ISO 22610.
Surgical drapes, gowns and clean air suits, used as medical devices, for patients, clinical staff and equipment - Test method to determine the resistance to wet bacterial penetration. Originally the Swedish standard HSS 8760019 was used as a basis for these measurements.

Toothed wheel transmission, free of slippage, load compensated drive system
The RULLA II test device is supplied with a fixed ratio toothed wheel transmission which works free of slippage and guarantees to reliably cover the complete test-surface during the preset time of 15 minutes. The load compensated drive system ensures a constant rotational speed of the agar plate holder of 60 ± 1 rpm over the entire range of 1 N to 3 N.

Programmable timer
The timer is preset by the factory to 15 minutes, i.e. the proper period necessary to perform the test according to the above standard. The digital timer is user-programmable to set other test-periods in the range from 1 second to 99 minutes and 59 seconds with an accuracy of ± 1 sec. The large LED display shows either the preset test period or while in operation shows the remaining run-time.

Housing, finger and turntable made of stainless steel
All essential parts of the housing are made of stainless steel, i.e. can readily be cleaned. The finger and the agar plate holder are made of stainless steel and can easily be removed for cleaning and disinfecting.

Optional accessories
Optionally available are the major accessories to perform the test. The support-cylinder is used during preparation for the test to stabilize the cover-/ carrier-/ reference material and assists in getting a smooth working surface of the materials inside the tenter frame. Not offered are additives to the agar and the strain of S. aureus.

A dynamometer attachment point is located in-line with the center of the finger for a spring balance to adjust the downward force of the finger onto the test material.

Both, the Swedish as well as the previous Schü tt Labortechnik GmbH manufactured apparatus made use of the slippage between the drive belt and the driven wheel to ensure an even coverage of the test material' s surface. Due to the increase in downward force of the finger from 1 N to 3 N ( as stipulated by EN ISO 22610) the slippage in these earlier devices becomes very difficult to control and is likely to render less reproducible results.